PRODUCT INFORMATION
Model # | E3641 |
---|---|
Separating Force | Typically 9 N to 14 N (2 lbf to 3 lbf) |
Gage Length | 3.0 mm to 12.5 mm (0.12 in. to 0.49 in.) |
Temperature Rating | Ambient to 540°C (1,000°F) |
For fracture mechanics tests in environmental chambers and forced exposure to heat. These capacitive sensor based clip-on gages (or COD gages) can be used up to 540 C (1000 F) with out any cooling. Length is 3 to 12.5 mm (and 0.2 to 0.5 in). Cable is triaxial ceramic fiber insulated cable 3 dt (1 m) plus 10 dt (2.5 m) extension cable. Output is user specified, +/-5 VDC or +/-10VDC typical. Linearity is <0.10% of full scale measuring range.
Features & Benefits
- Rugged design gives dynamic testing capability to 20 Hz or faster.
- Shart grooves per ASTM E1820, E813, and E399 for improved stability when mounting.
- Shipped fully calibrated with specified voltage output.
- 3603 signal conditioner included.
- Provides high level analog DC voltage output with very low noise.