The Luer taper is a standardized system of small-scale fluid fittings for making leak-free connections between sections on medical and laboratory instruments, including hypodermic syringe tips, catheters and infusion devices. ISO 594-1 is the standard specification and test method for the design of such Luer taper fittings. The two varieties of Luer Taper connections are typically known as Luer-Lock and Luer-Slip. The Luer-Lock standard connection can be tested according to ISO 594-2.
ISO 594-1 Separation Force Testing for Luers
ISO 594-1 provides test methods for testing the separation force of Luer fitting assemblies. For testing of separation force and unscrewing torque, take a look at ISO 594-2. Separation force is tested by applying a tensile axial force from a fatigue test machine. The testing standard asks for a tensile axial force to be applied progressively up to 25 N, at a rate of 10 N/s, and holding the force for at least 10 seconds.
For all tests in ISO 594-1 requiring an assembled Luer fitting, fittings are assembled using an axial torsion machine, applying a compressive force not greater than 27.5 N and a torque not greater than 0.1 Nm without exceeding 90 degrees rotation.
ISO 594-2 Separation Force and Unscrewing Torque Testing for Luer Tapers
ISO 594-2 provides test methods for testing the separation force of Luer fitting assemblies and unscrewing torque of Luer fitting assemblies. For testing of separation force without unscrewing torque, take a look at ISO 594-1.
Separation force is tested by applying a tensile axial force progressively up to 35 N, at a rate of 10 N/s, and holding the force for at least 10 seconds. Unscrewing torque is tested by applying a 0.02 Nm counter-clockwise torque, maintaining this torque for at least 10 seconds. For all tests in ISO 594-2 requiring an assembled Luer fitting, fittings are assembled using an axial torsion machine, applying a compressive force not greater than 27.5 N and a torque not greater than 0.12 Nm.
Equipment Required
By means of our testing expertise and modular product design, we will help find the solution that is right for you. Give our test engineers a call today for help configuring the best test machine and accessories according to your standard.
Recommended Test Machine
Force ratings up to 10 kN (2250 lbf) and torque ratings up to 100 Nm (885 in-lb)
Can apply axial and torsional loads simultaneously
Performs both static and slow-speed dynamic testing
All electric and oil-free operation
Force ratings up to 2.5 kN (560 lbf) and torque ratings up to 10 Nm (89 in-lb)
Can apply axial and torsional loads simultaneously
Axial frequencies up to 15 Hz and torsional frequencies up to 30 Hz
All electric and oil-free operation
Force ratings up to 50 kN (11,250 lbf) and torque ratings up to 200 Nm (1770 in-lb) Can apply axial and torsional loads simultaneously Performs both static and dynamic tests All electric and oil-free operation
Recommended Testing Accessories
Designed in accordance to ISO 594-1 Figure 3A
Includes certification
Optional handle available
>Designed in accordance to ISO 594-1 Figure 3B
Includes certification
Optional handle available
Designed in accordance to ISO 594-1 Figure 3C
Includes certification
Optional handle available
Designed in accordance to ISO 594-1 Figure 4
Includes certification
Designed in accordance to ISO 594-1 Figure 5
Includes certification
Designed in accordance to ISO 594-2 Figure 5
Includes certification
Designed in accordance to ISO 594-2 Figure 6
Includes certification
Designed in accordance to ISO 594-2 Figure 7
Includes certification
Designed in accordance to ISO 594-2 Figure 8
Includes certification